Suitability of capillary blood glucose analysis in patients receiving vasopressors
Myra F Ellis1, Kesi Benjamin, Morgan Cornell
1Duke University Hospital, Durham, NC, USA.
Background:
Glycemic control in critically ill patients decreases infection and mortality. Patients receiving vasopressors have altered peripheral perfusion, which may affect accuracy of capillary blood glucose values measured with point-of-care devices.
Objectives:
To compare capillary and arterial glucose values measured via point-of-care testing (POCT) with arterial glucose values measured via clinical chemistry laboratory testing (CCLT) in patients after cardiothoracic surgery. To determine if vasopressors or diminished peripheral perfusion influence the accuracy of POCT values.
Methods:
In a prospective, convenience sample of 50 adult postoperative cardiothoracic patients receiving insulin and vasopressors, 162 samples were obtained simultaneously from capillary and arterial sites during insulin infusion and tested via both POCT and CCLT. Clarke error grid analysis and ISO 15197 were used to analyze level of agreement. Two-way analysis of variance was used to analyze differences in glucose values with respect to vasopressor use and peripheral perfusion.
Results:
An unacceptable level of agreement was found between the capillary POCT results and arterial CCLT results (only 88.3% of values fell in zone A, or within the ISO 15197 tolerance bands). Arterial POCT results showed acceptable (94.4%) agreement with CCLT results. Vasopressor use had a significant effect on the accuracy of arterial blood glucose values (F=15.01; P<.001).
Conclusions:
Even when the more accurate POCT with arterial blood is used, blood glucose values are significantly less accurate in patients receiving more than 2 vasopressors than in patients receiving fewer vasopressors. CCLT may be safer for titrating insulin doses in these patients.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Hypoglycemia and Glucagon
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...