Outbreak of postoperative empyema caused by Serratia marcescens in a thoracic surgery unit
A Ulu-Kilic1, O Parkan, S Ersoy
1Department of Infectious Diseases, Faculty of Medicine, Erciyes University, Kayseri, Turkey.
Abstract:
An increase in the number of cases of postoperative empyema due to S. marcescens was recognized in the intensive care unit (ICU) of our Division of Thoracic Surgery between 3 and 19 March 2013. Pleural samples from patients and environmental samples from the operating room and ICU were obtained. A total of eight isolates (six from pleural fluid and two from portable suction devices in ICU) were identified as Serratia marcescens. All isolates were found to be identical by repetitive sequence-based polymerase chain reaction. This is the first report of an outbreak caused by S. marcescens related to a contaminated portable suction machine.
Insights
A Serratia marcescens outbreak in a thoracic surgery intensive care unit was linked to contaminated portable suction devices. This highlights the importance of environmental hygiene in preventing hospital-acquired infections.
Area of Science:
- Medical Microbiology
- Infectious Diseases
- Hospital Epidemiology
Background:
- Postoperative empyema cases caused by Serratia marcescens increased in a thoracic surgery intensive care unit (ICU).
- Serratia marcescens is an opportunistic pathogen often associated with hospital-acquired infections.
Purpose of the Study:
- To investigate the source of a Serratia marcescens outbreak in an ICU.
- To identify the link between clinical isolates and environmental samples.
Main Methods:
- Collection of pleural fluid samples from patients and environmental samples from the ICU and operating room.
- Identification of Serratia marcescens isolates.
- Repetitive sequence-based polymerase chain reaction (rep-PCR) for strain typing.
Main Results:
- Eight Serratia marcescens isolates were identified, with six from pleural fluid and two from portable suction devices.
- All eight isolates were genetically identical, confirmed by rep-PCR.
- This is the first reported outbreak of Serratia marcescens linked to contaminated portable suction machines.
Conclusions:
- Contaminated portable suction devices were identified as the source of the Serratia marcescens outbreak.
- Strict environmental cleaning protocols are crucial to prevent nosocomial infections.
- This outbreak underscores the need for vigilance in monitoring and controlling microbial contamination in healthcare settings.
Related Concept Videos
Pneumonia I: Introduction
Pneumonia I: Introduction
Risk Factors
Various factors influence the likelihood of developing pneumonia. Age plays a crucial role, with infants, children under two, and individuals over 65 at increased risk due to their...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Pneumonia III: Complications and Assessment
Esophageal Perforation-II: Clinical Manifestations and Management
Clinical Manifestations:
Atypical Pneumonia