Early follow-up of a custom non-fluted diaphyseal press-fit tumour prosthesis
Patrick W O'Donnell1, Anthony M Griffin, William C Eward
1Department of Orthopaedic Surgery, Markey Cancer Center, University of Kentucky, Lexington, KY, USA, patrick.odonnell@uky.edu.
Purpose:
The objective of this study was to evaluate the early results of a custom non-fluted diaphyseal press-fit stem for use with the global modular replacement system (GMRS) tumour prosthesis and the early complications associated with this implant.
Methods:
A total of 53 patients (54 implants) were identified from a prospective database where a custom non-fluted diaphyseal press-fit stem was used as part of the reconstruction of the limb. All patients had a minimum of 22 months of follow-up.
Results:
The rates of stem revision for any reason were calculated. The median follow-up was 36 months (range 22-85 months). Aseptic loosening was not observed in any patient.
Conclusions:
At early term follow-up, an uncemented non-fluted stem used with the GMRS tumour endoprosthesis provides a stable bone-prosthesis interface with no evidence of aseptic loosening.