Automated laboratory based X-ray beamline with multi-capillary sample chamber
S Purushothaman1, B L L E Gauthé, N J Brooks
1Department of Chemistry, Imperial College London, South Kensington Campus, London SW7 2AZ, United Kingdom.
Abstract:
An automated laboratory based X-ray beamline with a multi-capillary sample chamber capable of undertaking small angle X-ray scattering measurements on a maximum of 104 samples at a time as a function of temperature between 5 and 85 °C has been developed. The modular format of the system enables the user to simultaneously equilibrate samples at eight different temperatures with an accuracy of ±0.005 °C. This system couples a rotating anode generator and 2D optoelectronic detector with Franks X-ray optics, leading to typical exposure times of less than 5 min for lyotropic liquid crystalline samples. Beamline control including sample exchange and data acquisition has been fully automated via a custom designed LabVIEW framework.