Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Clinical response with adaptive CRT algorithm compared with CRT with echocardiography-optimized atrioventricular
Jagmeet P Singh1, William T Abraham, Eugene S Chung
1Massachusetts General Hospital Heart Center, Harvard Medical School, 55 Fruit St., Boston, MA 02114, USA.
Aims:
Adaptive cardiac resynchronization therapy (aCRT) is a novel algorithm for CRT pacing that provides automatic ambulatory selection between synchronized left ventricular (LV) or bi-ventricular (BiV) pacing and optimization of atrioventricular (AV) and inter-ventricular (VV) delays based on periodic measurement of intrinsic conduction. We aimed to compare the clinical response between aCRT and standard CRT in historical trials.
Methods And Results:
The treatment arm of the aCRT trial was compared with a pooled historical control (HC) derived from the CRT arms of four clinical trials (MIRACLE, MIRACLE ICD, PROSPECT, and InSync III Marquis) with respect to the proportion of patients who had an improved clinical composite score (CCS) at the 6-month follow-up. Patients in the HC underwent echocardiography-guided AV optimization after the implant. A propensity score model was used to adjust for 22 potential baseline confounders of the effect of CRT. Patients were stratified into quintiles according to the propensity score and the adjusted absolute treatment effect was obtained by averaging estimates across these quintiles. The propensity score model included 751 patients (aCRT: 266, historical trials: 485). The adjusted absolute difference in percent improved in CCS between the aCRT and HC arms was 11.9% [95% confidence interval (CI): 2.7-19.2%] favouring aCRT. The patients in the aCRT group were significantly more likely to have an improved CCS than the patients in the HC (odds ratio = 1.65, 95% CI: 1.1-2.5).
Conclusion:
The aCRT algorithm may be associated with additional improvement in clinical response compared with historical CRT with echocardiographic AV optimization.