Synchronized neonatal non-invasive ventilation-a pilot study: the graseby capsule with bi-level NCPAP
Debra J Stern1, Michael D Weisner, Sherry E Courtney
1Department of Pediatrics, Division of Neonatology, Stony Brook Medicine, Stony Brook, New York.
Objectives:
(1) To evaluate the Graseby capsule (GC) as a respiratory detection device when compared to respiratory inductance plethysmography (RIP); (2) to evaluate the response to the Graseby signal of a commercially available bi-level nasal CPAP device (BNCPAP) designed for use with the GC; and (3) to assess the performance of the GC/BNCPAP device when fitted on preterm infants.
Study Design:
The study consisted of four parts: (1) The response characteristics of the GC/BNCPAP were assessed without connection to an infant; (2) the respiratory detection of both GC and RIP were compared in six preterm infants (mean 1,242 g, range 900-1,530 g); (3) the GC/BNCPAP was connected in six preterm infants (mean 1,568 g, range 1,040-1,835 g), isolating the BNCPAP and the infant interaction by substituting an artificial "nose" for the infant and assessing performance using inspiratory times (Ti) of 0.1 through 0.5 sec with pressure levels of both 10/6 and 8/5 cmH2 O; and (4) the GC/BNCPAP was employed as a respiratory support device in six preterm infants (mean 1,189 g, range 785-1,795 g) using clinically required pressures and evaluating performance using Ti of 0.3, 0.4 and 0.5 sec.
Results:
(1) Within 26 ms of stimulation of the GC, the BNCPAP initiated air flow; however, the time to reach peak pressure was much longer; (2) the GC, when placed in the subxiphoid position, tracked the RIP signal nearly identically and occurred sooner; (3) a Ti of at least 0.3 sec was required to reach the desired high pressure setting; and (4) synchrony of the GC/BNCPAP occurred in 72-74% of infant breaths.
Conclusions:
The GC is a sensitive respiratory detection device; however, the GC/BNCPAP interface requires a minimum Ti of 0.3 sec and an adequate respiratory effort to achieve the desired pressure and to synchronously trigger the BNCPAP.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Suctioning the Nasopharyngeal Airway
Equipment Required
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...