Neonatal and pediatric manual hyperinflation: influence of oxygen flow on ventilation parameters
Pricila Mara Novais de Oliveira1, Armando Augusto Almeida-Junior, Celize Cruz Bresciani Almeida
1Department of Pediatrics, Faculty of Medical Sciences.
Background:
Although self-inflating bags are widely used for manual hyperinflation, they do not allow ventilation parameters, such as pressure or volume, to be set. We studied the ventilation performance of neonatal and pediatric self-inflating bags.
Methods:
We asked 22 physiotherapists to manually hyperinflate 2 lung models (neonatal and pediatric), using self-inflating bags from 3 manufactures (Hudson, Laerdal, and JG Moriya), with flows of 0, 5, 10, and 15 L/min. A pneumotachograph recorded tidal volume (V(T)), peak inspiratory pressure (PIP), peak inspiratory flow (PIF), peak expiratory flow (PEF), and inspiratory time.
Results:
The V(T), PIP, and inspiratory time delivered by the Hudson, Laerdal, and JG Moriya bags, in both neonatal and pediatric self-inflating bags, were significantly different (P < .001). The PEF and PIF delivered were different only when using the neonatal self-inflating bags (P < .001). The V(T), PIP, and PIF delivered with a flow of 0 L/min were lower than with 15 L/min (P < .05) with all the tested bags, in both the neonatal and pediatric sizes.
Conclusions:
The performance of the tested neonatal and pediatric bags varied by manufacturer and oxygen flow. There was an increase in VT, PIP, and PIF related to the increase of oxygen flow from 0 L/min to 15 L/min. The neonatal bags showed higher ventilation parameters variation than the pediatric bags.
Related Concept Videos
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...