Subject variability of shoulder abduction strength testing.

S P Magnusson1, G W Gleim, J A Nicholas

  • 1Nicholas Institute of Sports Medicine and Athletic Trauma, New York, New York 10021.

The American Journal of Sports Medicine
|July 1, 1990
PubMed
Summary

Shoulder abduction strength testing showed high reliability with both Cybex and hand-held dynamometers. Minimize measurement error by averaging specific repetitions for each device to determine clinical significance.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Physical Performance Tests in 8008 Competitive Youth Tennis Players-A Systematic Review and Meta-Analysis of Normative Values.

European journal of sport science·2025
Same author

Development and evaluation of a new semi-quantitative and morphometric scoring system for magnetic resonance imaging in adolescents with Osgood Schlatter Disease (The OSIS Score).

European journal of radiology·2025
Same author

Increased Risk of Tendon Injury Following Structured Care in Patients With Type 2 Diabetes: Post Hoc Analysis of a Large Randomized Controlled Trial With 19 Years of Follow-up.

Diabetes care·2024
Same author

A comprehensive MRI investigation to identify potential biomarkers of Osgood Schlatter disease in adolescents: A cross sectional study comparing Osgood Schlatter disease with controls.

Scandinavian journal of medicine & science in sports·2024
Same author

Determination of differences in ultrasound parameters for patellar tendons in males with unilateral patellar tendinopathy-An ancillary analysis of data from two randomized controlled trials.

Journal of clinical ultrasound : JCU·2024
Same author

Hormonal, immune, and oxidative stress responses to blood flow-restricted exercise.

Acta physiologica (Oxford, England)·2023

Area of Science:

  • Biomechanics
  • Kinesiology
  • Rehabilitation Science

Background:

  • Accurate measurement of shoulder abduction strength is crucial for assessing patient recovery and guiding rehabilitation.
  • Understanding the biological variability and measurement error associated with different strength testing devices is essential for clinical decision-making.

Purpose of the Study:

  • To quantify the normal biological variability in shoulder abduction strength testing using an isokinetic dynamometer (Cybex) and a hand-held dynamometer.
  • To determine the optimal method for minimizing interday variability for each testing modality.
  • To establish thresholds for measurement error to differentiate true strength changes from random variation.

Main Methods:

  • Nine healthy subjects underwent repeated shoulder abduction strength testing over five days.

Related Experiment Videos

  • One shoulder was tested using a Cybex isokinetic dynamometer at 60 deg/sec; the contralateral shoulder was tested with a hand-held dynamometer.
  • Six maximal voluntary contractions were performed per session, with sessions separated by 1-2 weeks.
  • Main Results:

    • High intraday and interday correlations were observed for both Cybex (0.88-0.996) and hand-held dynamometer (0.82-0.995) testing.
    • Interday variability was minimized by averaging the first three repetitions for Cybex (8.6%-19.2% SEM) and the last three for hand-held dynamometer (5.5%-10.8% SEM).
    • A significant decline in strength was noted for the hand-held dynamometer (last vs. first three reps), but not for the Cybex.

    Conclusions:

    • Both Cybex and hand-held dynamometers demonstrate high reliability for shoulder abduction strength testing.
    • Specific repetition averaging strategies (first three for Cybex, last three for hand-held) effectively reduce interday variability.
    • Strength changes below 19% for Cybex and 11% for hand-held dynamometers should be considered within the margin of measurement error and not clinically significant.