Validation of a specific prolylcarboxypeptidase activity assay and its suitability for plasma and serum measurements
Kaat Kehoe1, Robert Verkerk, Yani Sim
1Laboratory of Medical Biochemistry, Department of Pharmaceutical Sciences, University of Antwerp, 2610 Antwerp, Belgium.
Abstract:
Prolylcarboxypeptidase (PRCP, EC 3.4.16.2), a lysosomal carboxypeptidase, was discovered 45 years ago. However, research has been hampered by a lack of well-validated assays that are needed to measure low activities in biological samples. Two reversed-phase high-performance liquid chromatography (RP-HPLC) methods for quantifying PRCP activity in crude homogenates and plasma samples were optimized and validated. PRCP activity was determined by measuring the hydrolysis of N-benzyloxycarbonyl-l-proline (Z-Pro)-Phe. The enzymatically formed Z-Pro and Phe were measured independently under different HPLC conditions. The in-house methods showed good precision, linearity, accuracy, and specificity. Based on Michaelis-Menten constants, Z-Pro-Phe was chosen over Z-Pro-Ala as the substrate of preference. Cross-reactivity studies with dipeptidyl peptidases (DPPs) 2, 4, and 9 and prolyl oligopeptidase (PREP) confirmed the specificity of the PRCP activity assay. The average PRCP activity in plasma and serum of 32 healthy individuals was found to be 0.65 ± 0.02 and 0.72 ± 0.03 U/L, respectively. Both methods can be used to measure PRCP activity specifically in different biological samples and are well suited to evaluate PRCP inhibitors. These well-validated methods are valuable tools for studying PRCP's role in cardiovascular diseases, stroke, inflammation, and metabolic syndrome.