Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
Dosimetric evaluation of dental implant planning examinations with cone-beam computed tomography
Marcos E A Andrade1, Helen J Khoury, João B S Nascimento Neto
1Departamento de Energia Nuclear, Universidade Federal de Pernambuco-DEN/UFPE, Av. Prof. Luiz Freire, 1000, Recife, PE 50740-540, Brazil.
Abstract:
The aim of this work was to perform a dosimetric evaluation of two cone-beam computed tomography scanners. The study was performed with two scanners: i-CAT classic and PreXion 3D. Air kerma-area product (PKA) was measured for all full-arch dental implant planning protocols. Surface air kerma was also estimated at the region of the eyes, salivary glands and thyroid using thermoluminescence dosemeters positioned on an anthropomorphic phantom. The PKA values for the i-CAT classic ranged from 24 to 180 µGy m(2) and, for the PreXion, from 70 to 138 µGy m(2). The large variation of these values was mainly caused by acquisition time and field of views. The surface air kerma values were from 0.08 to 3.39 mGy at the eyes, 0.50 to 3.96 mGy at the parotids, 0.11 to 2.95 mGy at the submandibular glands and 0.05 to 1.32 mGy at the thyroid. These values are comparable with those found in the literature.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography