Design and demonstration of a battery-less fluid warmer for combat
Background:
Prehospital battlefield hypothermia remains an issue, with cold fluid resuscitation likely being a significant contributor. Currently, no prehospital battlefield technology exists to warm intravenous resuscitation fluids. Existing commercial fluid-warming technologies are either inadequate or unreliable or have an unacceptable weight and size, making them inappropriate for the austere combat environment. We propose the creation of a battery-less, flameless, portable, lowweight, small, chemically powered fluid warmer for the battlefield.
Methods:
A magnesium-based exothermic chemical reaction was used as the sole heating source. A low-weight, small insulated container was created to contain the reaction. The chemical reaction was manipulated to sustain fluid heating as long as required.
Results:
The exothermic reaction was used to boil a Fluorinert ™ liquid within an insulated container that heats resuscitation fluid passing through the heat exchanger. A working prototype device, 9 inches in length and 4 inches in diameter, was engineered and tested. Warming was maintained over a variety of clinically relevant flow rates.
Conclusion:
A chemically based, safe, battery-less, flameless, lightweight fluid warmer was created. This technology could represent a significant remote capability currently unavailable on the battlefield.
More Related Videos
13:59Reduced-gravity Environment Hardware Demonstrations of a Prototype Miniaturized Flow Cytometer and Companion Microfluidic Mixing Technology
Published on: November 13, 2014
08:50Fabrication and Characterization of a Conformal Skin-like Electronic System for Quantitative, Cutaneous Wound Management
Published on: September 2, 2015
Related Concept Videos
Design Example: Flow Through a Fire Extinguisher
The key to understanding how the...
DC Battery