An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Stress-strain distribution at bone-implant interface of two splinted overdenture systems using 3D finite element
1Prosthodontics, Prosthodontic Department, Faculty of Dentistry, Qassim University. Member, Dental Implant Committee, Dental Research Center Board Member, College of Dentistry, Qassim University, Qassim, Kingdom of Saudi Arabia.
Purpose:
This study was accomplished to assess the biomechanical state of different retaining methods of bar implant-overdenture.
Materials And Methods:
Two 3D finite element models were designed. The first model included implant overdenture retained by Hader-clip attachment, while the second model included two extracoronal resilient attachment (ERA) studs added distally to Hader splint bar. A non-linear frictional contact type was assumed between overdentures and mucosa to represent sliding and rotational movements among different attachment components. A 200 N was applied at the molar region unilaterally and perpendicular to the occlusal plane. Additionally, the mandible was restrained at their ramus ends. The maximum equivalent stress and strain (von Mises) were recorded and analyzed at the bone-implant interface level.
Results:
The values of von Mises stress and strain of the first model at bone-implant interface were higher than their counterparts of the second model. Stress concentration and high value of strain were recognized surrounding implant of the unloaded side in both models.
Conclusion:
There were different patterns of stress-strain distribution at bone-implant interface between the studied attachment designs. Hader bar-clip attachment showed better biomechanical behavior than adding ERA studs distal to hader bar.