Dosimetry for Cell Irradiation using Orthovoltage (40-300 kV) X-Ray Facilities
Published on: February 20, 2021
Dose distribution response in HDRB measured with EBT2 and compared with PLATO SYSTEM
L Hernández-Ruiz1, J O Hernández-Oviedo, D Ruesga-Vazquez
1Centro de Investigación en Ciencia Aplicada y Tecnología Avanzada-Legaria del Instituto Politécnico Nacional, Av. Legaria 694, 11500 México City, Mexico.
Abstract:
Dose distribution of a High Dose Rate Brachytherapy (BHDR) oncological treatment with (192)Ir was measured using a Gafchromic EBT2 film. The film calibration was performed with a (60)Co unit and a LINAC of 6 mV and 18 mV. Gafchromic behavior of a dosimeter varies in respect of energy. Experimental results of dose distribution match with those planned in the PLATO commercial system, they also show that there is a difference of 2.11% between the planning system and isodoses measured.