FloTrac/Vigileo(TM) (third generation) and MostCare(®)/PRAM versus echocardiography for cardiac output estimation in
Stefano Romagnoli1, Zaccaria Ricci, Salvatore M Romano
1Department of Heart and Vessels, Cardio-Thoracic and Vascular Anaesthesia and Intensive Care, Azienda Ospedaliero-Universitaria Careggi, Florence, Italy.
Objective:
To compare the FloTrac/Vigileo(TM) cardiac output (COFT/V) and the MostCare(®)/PRAM cardiac output (COMC/P) versus transthoracic echocardiographic cardiac output estimation (reference method; CO(ECHO)).
Design:
Prospective observational study.
Setting:
Single center, Cardio-Thoracic and Vascular Surgery/Intensive Care Unit.
Participants:
Patients undergoing elective vascular surgery.
Interventions:
Cardiac output measurement with two pulse contour methods: the FloTrac/Vigileo(TM) and the MostCare(®)/PRAM before (T1) and after (T2) fluid loading versus echocardiography (reference method).
Measurements And Main Results:
One hundred fifty-six CO measurements were performed in 26 patients. The data showed poor agreement between CO(ECHO) and CO(FT/V): r(2) = 0.29 (T1) and 0.27 (T2); bias -0.37 (T1) and -0.40 (T2) L/min; limits of agreement from -3.10 to 2.42 (T1) and from -3.0 to 2.2 (T2) L/min. The percentage error was 51.7% (T1) and 49.3% (T2). Conversely, COMC/P resulted in agreement with echocardiography: r(2) = 0.76 (T1) and 0.80 (T2); bias -0.01 (T1) and -0.06 (T2) L/min; limits of agreement from -1.13 to 1.11 (T1) and from -0.90 to 0.80 (T2) L/min, with a PE of 22.4% (T1) and of 17.0% (T2).
Conclusions:
In patients undergoing vascular surgery, the FloTrac/Vigileo(TM) did not demonstrate that it was a reliable system for CO monitoring when compared with echocardiography-derived CO. However, MostCare(®)/PRAM was shown to estimate CO with a good level of agreement with echocardiographic measures.
Related Concept Videos
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...