Objective:
Experience of continuous monitoring and control of patient doses in CT in Dubai Hospital over a period of approximately 4 years (January 2008 through August 2011) is presented.
Materials And Methods:
Dose measurements-in particular, weighted and volumetric CT dose index, dose-length product (DLP), and estimated effective dose-were regularly monitored using head (16 cm diameter) and body (32 cm diameter) CT phantoms. Patient radiation dose indexes were manually recorded during 2008 for common CT examinations: head, chest, and abdomen and pelvis scans. In 2009-2011, these CT dose data were recorded within the radiology information system and the PACS. Dose reduction actions were taken while maintaining a watch on image quality. The effects of these factors were monitored through change in average DLP on a monthly basis and third quartile annually. Adapted diagnostic reference levels were used for comparison.
Results:
The reduction in adult dose indexes in 2010 as compared with 2008 was 52%, 16.4%, and 34.8% for head, chest, and abdomen and pelvis examinations, respectively. For the pediatric group, the reduction was 45.23%, 39.6%, and 43.34% for head, chest, and abdomen and pelvis examinations, respectively.
Conclusion:
Substantial reduction in DLP for common examinations of adults and children is shown through a program of continuous monitoring and action. The results indicate the need to introduce local diagnostic reference levels to substitute for the adapted ones.
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Therapeutic Drug Monitoring: Overview and Classification