Related Experiment Video
Updated: May 7, 2026

Merging Ion Concentration Polarization between Juxtaposed Ion Exchange Membranes to Block the Propagation of the Polarization Zone
Published on: February 23, 2017
Ion-Selective Electrodes for the Potentiometric Determination of Pramoxine HCl Using Different Ionophores
Abstract:
Four novel pramoxine HCl (PAM) selective electrodes were investigated with 2-nitrophenyl octylether as a plasticiser in a polymeric matrix of polyvinyl chloride (PVC). Sensor 1 was fabricated using sodium-tetraphenylborate (TPB) as an anionic exchanger without incorporation of an ionophore. Sensor 2 used 2-hydroxy propyl -cyclodextrin as an ionophore, while sensors 3 and 4 were constructed using 4-sulfocalix-6-arene and 4-sulfocalix-8-arene respectively as ionophores. Linear responses of PAM within the concentration ranges of 1.0 × 10-4 to 1.0 × 10-2 mol L-1 and 1.0 × 10-5 to 1.0 × 10-2 mol L-1 were obtained using sensors 1 and 2, respectively and 1.0 × 10-6 to 1.0 × 10-2 mol L-1 were obtained using sensors 3 and 4. Nernstian slopes of 50.4 ± 0.6, 54.3 ± 0.8, 56.3 ± 0.3 and 59.1 ± 0.5 mV/decade over the pH range of 3.0-6.0 were observed. The selectivity coefficients of the developed sensors indicated excellent selectivity for PAM. The utility of 2-hydroxy- propylcyclodextrin (2HP-β-CD) and 4-sulfocalix [6,8] arene (SC 6, 8) as ionophores had a significant influence on increasing the membrane sensitivity and selectivity of sensors 2, 3 and 4 compared to sensor 1. The proposed sensors displayed useful analytical characteristics for the determination of PAM in bulk powder, pharmaceutical formulation, and in biological fluid. Validation of the method showed the suitability of the proposed electrodes for the use in the quality control assessment of the drug. Furthermore, statistical comparison between the results obtained by the proposed method and the official method of the drug was performed and no significant difference was found.
More Related Videos
07:38Functional Characterization of Na+/H+ Exchangers of Intracellular Compartments Using Proton-killing Selection to Express Them at the Plasma Membrane
Published on: March 30, 2015
08:39Whole-cell Patch-clamp Recordings for Electrophysiological Determination of Ion Selectivity in Channelrhodopsins
Published on: May 22, 2017
Related Concept Videos
Potentiometry: Membrane Electrodes
Ion-Exchange Chromatography
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometry: Overview
Potentiometric Titration: Overview
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...