In situ Compressive Loading and Correlative Noninvasive Imaging of the Bone-periodontal Ligament-tooth Fibrous Joint
Published on: March 7, 2014
Biomechanical strength analysis of mini anchors for the temporomandibular joint
Andrezza Lauria1, Danillo Costa Rodrigues, Fábio Ricardo Loureiro Sato
1Department of Oral and Maxillofacial Surgery, School of Dentistry of Piracicaba, State University of Campinas (UNICAMP), Avenida Limeira, 901, Caixa Postal 52, #191, Areião, 13414-903, Piracicaba, São Paulo, Brazil, andrezzalauria@gmail.com.
Introduction:
The usage of mini anchors for the treatment of internal derangements of the temporomandibular joint (TMJ) is considered an innovative treatment, and it has presented excellent clinical results in the stabilization of the articular disc. This study aimed to evaluate, through mechanical tensile testing, the resistance of mini anchors for TMJ articular disc repositioning.
Materials And Methods:
Ten thermoplastic polymer poly-ether-ether-ketone (PEEK) mini anchors were tested in artificial polyurethane blocks with a pullout test in a mechanical Instron™ machine, model 4411. The mini anchors were pulled out until failure and the force and displacement were recorded for each specimen. A small standard deviation was noted, which indicated reproducibility.
Results:
The findings indicate that Cillen™ mini anchors presented a mean force at failure up to 46.1 N with similar values in the scientific literature. Nevertheless, further studies should be conducted for a comparison of PEEK mini anchors with other commercially available anchors, in addition to studies related to PEEK, a scarcely known material in the dentistry field.