Catherine Dicaire1, Jean-Nicholas Mess, Milton Furtado
1Algorithme Pharma, Inc., 575 Armand-Frappier, Laval, QC, Canada.
Background:
Matrix effects are one of the major drawbacks of ESI-MS/MS. It is majorly caused by lipids in plasma, which can be overcome by using different extraction techniques.
Results:
In this investigation, a major matrix effect was observed in samples containing a co-administered drug. Unknown compounds appeared over time in the human plasma samples spiked with co-administered drug creating major ion suppression. The changes in matrix integrity were associated with the organic solvent content in the plasma samples.
Conclusion:
The amount and type of organic solvent added to human plasma along with the storage conditions must be carefully determined during method development in order not to impact quantitation.