Trabecular Meshwork Response to Pressure Elevation in the Living Human Eye
Published on: June 20, 2015
Does rebound tonometry probe misalignment modify intraocular pressure measurements in human eyes?
Ian G Beasley1, Deborah S Laughton, Benjamin J Coldrick
1Ophthalmic Research Group, Life and Health Sciences, Aston University, Birmingham B4 7ET, UK.
Journal of Ophthalmology
|September 28, 2013
Summary
The iCare rebound tonometer overestimates intraocular pressure (IOP) compared to Goldmann applanation tonometry. However, IOP measurements using the iCare tonometer are generally unaffected by minor positional misalignments in healthy eyes.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Medical Devices
Background:
- Intraocular pressure (IOP) is a key metric in diagnosing and managing glaucoma.
- Rebound tonometry offers a portable and non-invasive method for IOP measurement.
- Understanding the accuracy and limitations of rebound tonometers, such as the iCare device, is crucial for clinical application.
Purpose of the Study:
- To evaluate the impact of positional misalignments on IOP measurements obtained with the iCare rebound tonometer.
- To compare iCare tonometer readings against Goldmann applanation tonometer (GAT) measurements under various conditions.
Main Methods:
- IOP was measured in 36 healthy subjects using the iCare rebound tonometer at the central corneal apex (CC) and compared with GAT.
- The iCare tonometer was also used to take readings at lateral displacements (2 mm nasal/temporal) and angular deviations (5° and 10° nasal/temporal) from the visual axis.
- A bespoke rig was utilized to control positional accuracy during measurements.
Main Results:
- The iCare tonometer consistently overestimated IOP compared to GAT (17.4 ± 3.6 mmHg vs. 14.7 ± 2.5 mmHg, P < 0.001).
- This overestimation increased with higher average IOPs.
- IOP measurements were not significantly affected by lateral displacements.
- Marginal underestimation of IOP was observed with angular deviations, reaching statistical significance only at 10° nasally (<1 mmHg).
Conclusions:
- The iCare rebound tonometer overestimates IOP relative to GAT, a known limitation.
- Despite this, the iCare tonometer demonstrates robustness against positional misalignments in healthy individuals.
- Minor angular deviations, such as 10° nasally, result in statistically, but not clinically, significant underestimations.