Accuracy of a home-based device for giving an early estimate of pregnancy duration compared with reference methods
Sarah R Johnson1, Sonya Godbert, Paul Perry
1SPD Development Company Limited, Bedford, United Kingdom.
Objective:
To assess a home pregnancy test's accuracy to concurrently detect pregnancy and determine pregnancy duration.
Design:
Multicenter, prospective study.
Setting:
Study sites in the United States.
Patient(S):
Women actively attempting to conceive who have menstrual bleeds (18-45 years).
Intervention(S):
Volunteers collected early morning urine samples (three or fewer menstrual cycles). Pregnant volunteers underwent ultrasound dating scans. Ovulation day (LH surge +1 day) during pregnancy-resulting cycles was determined by quantitative measurement of LH. Random urine samples were tested with the hCG-measuring pregnancy test from 4 days before the expected period until 4 weeks later.
Main Outcome Measure(S):
A home pregnancy test's accuracy in determining pregnancy duration compared with ultrasound and ovulation day.
Result(S):
Agreement between pregnancy test results and time since ovulation was 93% (confidence interval [CI], 91.5-94.4). Agreement with ultrasound was dependent on the formula: there was 99% agreement when calculated with adjustment for Hadlock formula bias (Pexsters; CI, 98.2-99.4) or using a nonbias formula (Wu; CI, 98.6-99.6), when ultrasound error was accommodated. Agreement was lower when bias/measurement errors were not accounted for (Wu, 86%, CI, 83.9-88; Hadlock, 80.8, CI, 78.2-83.3).
Conclusion(S):
This home pregnancy test provides an accurate estimation of pregnancy duration in weeks categories, 1-2, 2-3, 3+ weeks since ovulation, thereby showing utility in dating pregnancy.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...