Inaccuracy of estimated resting oxygen uptake in the clinical setting
Nikhil Narang1, Jennifer T Thibodeau, Benjamin D Levine
1Department of Medicine, University of Chicago, Chicago, IL (N.N.); Department of Internal Medicine (J.T.T., B.D.L., M.O.G., A.T.T., J.A.d.L.,D.K.M.) and Cardiovascular Division (J.T.T., B.D.L., M.O.G., A.T.T., J.A.d.L., D.K.M.), the University of Texas Southwestern Medical Center, Dallas, TX; Institute of Exercise and Environmental Medicine at the Texas Health Presbyterian Hospital, Dallas, TX (B.D.L.); the Department of Clinical Sciences at the University of Texas Southwestern Medical Center, Dallas, TX (C.R.A., D.K.M.); Cardiology Division, University of Texas Health Science Center at San Antonio, San Antonio, TX (R.A.L.); and the Cardiovascular Division, Oregon Health and Science University, Portland, OR (J.E.C.).
Background:
The Fick principle (cardiac output = oxygen uptake ( O2)/systemic arterio-venous oxygen difference) is used to determine cardiac output in numerous clinical situations. However, estimated rather than measured O2 is commonly used because of complexities of the measurement, though the accuracy of estimation remains uncertain in contemporary clinical practice.
Methods And Results:
From 1996 to 2005, resting O2 was measured via the Douglas bag technique in adult patients undergoing right heart catheterization. Resting O2 was estimated by each of 3 published formulae. Agreement between measured and estimated O2 was assessed overall, and across strata of body mass index, sex, and age. The study included 535 patients, with mean age 55 yrs, mean body mass index 28.4 kg/m2; 53% women; 64% non-white. Mean (±standard deviation) measured O2 was 241 ± 57 ml/min. Measured O2 differed significantly from values derived from all 3 formulae, with median (interquartile range) absolute differences of 28.4 (13.1, 50.2) ml/min, 37.7 (19.4, 63.3) ml/min, and 31.7 (14.4, 54.5) ml/min, for the formulae of Dehmer, LaFarge, and Bergstra, respectively (P<0.0001 for each). The measured and estimated values differed by >25% in 17% to 25% of patients depending on the formula used. Median absolute differences were greater in severely obese patients (body mass index > 40 kg/m2), but were not affected by sex or age.
Conclusions:
Estimates of resting O2 derived from conventional formulae are inaccurate, especially in severely obese individuals. When accurate hemodynamic assessment is important for clinical decision-making, O2 should be directly measured.
More Related Videos
04:20Integration of Brain Tissue Saturation Monitoring in Cardiopulmonary Exercise Testing in Patients with Heart Failure
Published on: October 1, 2019
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Respiratory Volumes and Capacities I
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Errors occurring during blood pressure monitoring
Several factors...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...