Note: A self-calibrating electrometer for atmospheric charge measurements from a balloon platform
1Department of Meteorology, University of Reading, Reading, Berkshire RG6 6BB, United Kingdom.
The Review of Scientific Instruments
|October 5, 2013
Summary
A new disposable sensor measures electric charge in non-thunderstorm clouds. This instrument detected charge layers up to 40 pC m(-3) on stratocumulus cloud edges, improving atmospheric measurements.
Area of Science:
- Atmospheric Science
- Cloud Physics
- Aerosol Science
Background:
- Charged aerosol particles and water droplets are prevalent in the lower atmosphere.
- These charged particles can influence interactions within small cloud droplets.
- Understanding charge distribution is crucial for cloud dynamics and atmospheric electricity.
Purpose of the Study:
- To introduce a novel, small, disposable sensor for measuring electric charge in non-thunderstorm clouds.
- To improve upon previous sensor designs for atmospheric charge measurement.
- To provide a reliable method for in-situ charge detection on balloon platforms.
Main Methods:
- Development of a disposable sensor utilizing a self-calibrating current measurement technique.
- Integration of the sensor onto a free balloon platform alongside a standard meteorological radiosonde.
- Testing the sensor's performance across a temperature range of 5 °C to -60 °C, with stability within 5 fA.
Main Results:
- The sensor accurately measures currents ranging from 2 fA to 15 pA.
- Demonstrated stability of the sensor within 5 fA over a wide temperature range.
- Detected significant charge layers, up to 40 pC m(-3), at the edges of a stratocumulus cloud during a balloon flight.
Conclusions:
- The developed sensor offers a practical and improved method for measuring atmospheric charge in non-thunderstorm clouds.
- The detection of substantial charge layers highlights the importance of charge in stratocumulus cloud formations.
- This technology advances the capability for in-situ atmospheric electricity research.
Related Concept Videos
Galvanometer
Common devices, including car instrument panels, battery chargers, and inexpensive electrical instruments, measure potential difference (voltage), current, or resistance using a d'Arsonval galvanometer. This electromechanical instrument is also known as a moving coil galvanometer.
The galvanometer consists of two concave-shaped permanent magnets, providing a uniform radial magnetic field in the annular region. In the center, a pivoted coil of fine copper wire is placed in the uniform magnetic...
The galvanometer consists of two concave-shaped permanent magnets, providing a uniform radial magnetic field in the annular region. In the center, a pivoted coil of fine copper wire is placed in the uniform magnetic...
Instrument Calibration
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Potentiometer
Voltage and current measurements using a standard voltmeter and ammeter alter the circuit being measured either by drawing or resisting the current flow, which introduces uncertainties in the measurements. Null measurements balance the voltages so that no current flows through the measuring device and, therefore, no alterations occur in the measured circuit.
Suppose the emf of a battery needs to be measured. If the battery is directly connected to a standard voltmeter, the measured quantity is...
Suppose the emf of a battery needs to be measured. If the battery is directly connected to a standard voltmeter, the measured quantity is...
Coulometry: Overview
Coulometry is one of the rapid, most accurate, and precise analytical techniques that determine the quantity of an analyte by measuring the electrical charge needed for its complete electrolysis without using any analytical standards. The total charge passed during electrolysis correlates with the analyte amount by Faraday's laws of electrolysis. For accurate coulometric measurements, a charge equal to Faraday's constant multiplied by the number of electrons involved in the relevant...
Gravimetry: Overview
Gravimetric analysis is a quantitative method where the analyte is isolated and weighed directly or after conversion into a substance of known composition. Gravimetric analysis can be classified as precipitation, electrogravimetry, volatilization, and particulate gravimetry, based on the method used to isolate the analyte.
In precipitation gravimetry, the analyte is converted into a precipitate and weighed. For example, the silver content in a sample can be estimated by precipitating and...
In precipitation gravimetry, the analyte is converted into a precipitate and weighed. For example, the silver content in a sample can be estimated by precipitating and...
Amperometry: Overview
Amperometry is a technique commonly used to measure the concentration of specific analytes in a solution by monitoring the electric current generated during an electrochemical reaction. It involves applying a constant potential between a working electrode and a reference electrode to measure the resulting current, which is proportional to the concentration of the analyte. The Clark oxygen electrode operates based on this principle of amperometry. It consists of a cathode and an anode enclosed...