Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Simultaneous 99mTc-MDP/123I-MIBG tumor imaging using SPECT-CT: phantom and constructed patient studies
Yothin Rakvongthai1, Georges El Fakhri, Ruth Lim
1Center for Advanced Medical Imaging Sciences, Division of Nuclear Medicine and Molecular Imaging, Massachusetts General Hospital, 55 Fruit Street, Boston, Massachusetts 02114 and Department of Radiology, Harvard Medical School, 25 Shattuck Street, Boston, Massachusetts 02115.
Purpose:
Authors' goal is to evaluate the performance of simultaneous (99m)Tc-MDP/(123)I-MIBG tumor imaging with fast Monte-Carlo (MC) based joint iterative reconstruction as compared to sequential (99m)Tc-MDP and (123)I-MIBG tumor imaging.
Methods:
Noise-free (99m)Tc and (123)I SPECT projections were acquired separately using an anthropomorphic torso phantom modified to include a fillable tube around the lungs to mimic ribs. Additionally, (99m)Tc and (123)I projections were acquired separately using a 1-cm spherical "tumor" placed at various distances from one detector. Tumor-present data were generated by adding tumor projections to the torso phantom data, which were scaled to the total counts in typical clinical studies. Twenty-five noise realizations were generated by adding Poisson noise to the projection data for each radionuclide. Dual-radionuclide data were synthesized by summing the (99m)Tc and (123)I projections. Image reconstruction was performed using: (1) SR-OSEM, ordered subset expectation maximization (OSEM) without scatter correction (SC) using single-radionuclide (SR) data; (2) SR-MC-OSEM, OSEM with a fast MC-based SC using SR data; (3) DR-OSEM, OSEM without SC using dual-radionuclide (DR) data; and (4) DR-MC-JOSEM, joint OSEM with a fast MC-based SC using DR data. Ten (99m)Tc-MDP and ten (123)I-MIBG data sets, which had tumors mathematically inserted, were also used to evaluate the performance of authors' approach. For the phantom study, relative bias and relative standard deviation of tumor uptake were computed for each tumor using the tumor uptake in the noise-free single-radionuclide images, which were reconstructed by SR-MC-OSEM, as the gold standard. For both the phantom and constructed patient studies, mean contrast and standard deviation of contrast were computed for each tumor for both the single- and dual-radionuclide images. Additionally, contrast recovery was computed as the ratio between mean contrast and the mean contrast for SR-MC-OSEM.
Results:
For the phantom study, DR-MC-JOSEM yielded 2.7% on average relative bias of tumor uptake using the images, which were reconstructed from the noise-free SR data with SR-MC-OSEM, as the gold-standard. For both the phantom and constructed patient studies, DR-MC-JOSEM yielded 94.7% and 95.2% tumor contrast recovery on average using SR-MC-OSEM as the reference, in the phantom and constructed patient studies, respectively. DR-MC-JOSEM yielded comparable relative standard deviation of bias and standard deviation of contrast to SR-MC-OSEM.
Conclusions:
Simultaneous (99m)Tc-MDP/(123)I-MIBG tumor imaging using authors' dual-radionuclide reconstruction approach yielded comparable image quality to sequential (99m)Tc-MDP and (123)I-MIBG imaging. For patients who need to undergo both scans, authors' approach offers perfectly registered dual-tracer images under identical conditions without compromising image quality, and reduces the imaging cost while increasing patient throughput.
More Related Videos
07:54Implantation and Monitoring by PET/CT of an Orthotopic Model of Human Pleural Mesothelioma in Athymic Mice
Published on: December 21, 2019
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...