Background:
Flexible cystoscopy is used in urological outpatient departments for diagnostic cystoscopy of bladder cancer and requires a high-level disinfection between each patient. The purpose of this study was to make a microbiological post disinfection efficacy assessment of flexible cystoscopes (FC) using disposable sterile endosheaths.
Methods:
One hundred endosheaths underwent a leak-test for barrier integrity after cystoscopy. Microbiological samples from these cystoscopies were obtained; after removal of the endosheath, and after cleaning the scope with a detergent cloth, rinsing with tap water followed by 70% ethanol disinfection and subsequent drying. The number of colony forming units (cfu) from the samples was counted after 72 hours and then divided in three categories, Clean FC (<5 cfu/sample), Critical FC (5-50 cfu/sample) and High-risk FC (>50 cfu/sample). The result was compared with data of 10 years continuous control sampling recorded in the Copenhagen Clean-Endoscope Quality Control Database (CCQCD) and analyzed with a Chi-square test for homogeneity.
Results:
All 100 endosheaths passed the leak-test. All samples showed a Clean FC and low means of cfu. A query to the CCQCD, showed that 99.8% (1264/1267) of all FC with a built-in work-channel reprocessed in a WD were clean before use.
Conclusion:
The reprocessing of FC using endosheaths, as preformed in this study, provides a patient-ready procedure. The results display a reprocessing procedure with low risk of pathogen transmission, high patient safety and a valid alternative to the recommended high-level disinfection procedure of FC. However, the general impression was that sheaths slightly reduced vision and resulted in some patient discomfort.
More Related Videos
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Hand hygiene
Hand washing...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...