Abstract:
An experimental study was performed using a custom-built air-powered needle-free injector to investigate the various injector parameters governing the dynamics of jet injection. A parametric study using five different nozzle sizes at driver pressure ranging from 4 to 8 bar was carried out. The fluid stagnation pressure of the liquid jet was determined using a Honeywell force sensor. Performance plots as a function of various parameters were constructed. It was determined that as the driver pressure increased both the peak and average stagnation pressure increased almost linearly within the operating range considered. Varying the injection nozzle diameter, whilst keeping the driver pressure constant did not have any significant impact on the peak or average stagnation pressure. The chamber length was also varied and no significant influence was found on peak or average stagnation pressure.