Related Experiment Video
Updated: May 7, 2026

A Step by Step Protocol for Subretinal Surgery in Rabbits
Published on: September 13, 2016
Safety assessment of semichronic suprachoroidal electrical stimulation to rabbit retina
Abstract:
Confirming safety of chronic electrical stimulation is of prime importance for the practical use of visual prostheses. Here we applied electrical stimulation to eyes of freely-moving rabbits eight hour per day for one month. Examinations including fundus photo, optical coherence tomography (OCT), electrically evoked potentials (EEPs) were performed before and after one-month stimulation to detect tissue damage. No adverse effect caused by electrical stimulation was observed in electrophysiological and histological evaluation. We also found that there was no sign of morphological and electrochemical degradation of stimulating electrodes.
More Related Videos
12:01Techniques for Processing Eyes Implanted With a Retinal Prosthesis for Localized Histopathological Analysis
Published on: August 2, 2013
11:03Subretinal Transplantation of Human Embryonic Stem Cell Derived-retinal Pigment Epithelial Cells into a Large-eyed Model of Geographic Atrophy
Published on: January 22, 2018