Comparison of aneroid and oscillometric blood pressure measurements in children

Sigridur B Eliasdottir1, Sandra D Steinthorsdottir, Olafur S Indridason

  • 1Faculty of Medicine, School of Health Sciences, University of Iceland, Reykjavik, Iceland.

Insights

This study found minimal differences in systolic blood pressure (BP) between aneroid and oscillometric devices in children. However, diastolic BP was lower with oscillometric monitors, highlighting the need for multiple measurements.

Area of Science:

  • Pediatrics
  • Cardiovascular Health
  • Medical Devices

Background:

  • Accurate blood pressure (BP) measurement is crucial for diagnosing and managing hypertension in children.
  • Limited comparative data exist for aneroid and oscillometric BP devices in pediatric populations.
  • Understanding device-specific measurement differences is essential for consistent clinical practice.

Purpose of the Study:

  • To compare blood pressure readings obtained from oscillometric and aneroid BP monitors in 9- to 10-year-old children.
  • To investigate potential discrepancies in systolic and diastolic BP measurements between the two device types.
  • To inform best practices for pediatric BP monitoring.

Main Methods:

  • A total of 979 children aged 9-10 years participated in the study.
  • Participants were randomized into two groups: Group O (oscillometric then aneroid) and Group A (aneroid then oscillometric).
  • Two BP readings were taken with each device type sequentially within each group.

Main Results:

  • No significant difference was observed in mean systolic BP between oscillometric and aneroid devices (111.5±8.6 vs 111.3±8.1 mm Hg; P=.39).
  • Mean diastolic BP was significantly lower when measured with the oscillometric monitor compared to the aneroid device (61.5±8.0 vs 64.5±6.8 mm Hg; P<.001).
  • A downward trend in BP readings was noted with consecutive measurements, and limited agreement between devices was found.

Conclusions:

  • While systolic BP measurements showed good agreement, diastolic BP readings differed between oscillometric and aneroid devices in children.
  • The observed trend of decreasing BP with repeated measurements suggests the need for caution in diagnosis.
  • Multiple BP measurements using either device are recommended prior to diagnosing elevated BP or hypertension in pediatric patients.

Related Concept Videos

Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.4K
Sites for measuring blood pressure01:21

Sites for measuring blood pressure

Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
3.9K
Assessment of blood pressure in brachial artery(two-step method)01:23

Assessment of blood pressure in brachial artery(two-step method)

Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
2.1K