Laminated paper-based analytical devices (LPAD) with origami-enabled chemiluminescence immunoassay for cotinine
Wei Liu1, Christopher L Cassano, Xin Xu
1Interdisciplinary Microsystems Group, Department of Mechanical and Aerospace Engineering, University of Florida , P.O. Box 116250, Gainesville, Florida 32611, United States.
Abstract:
Laminated paper-based analytical devices (LPAD) with origami-enabled chemiluminescence immunoassay have been developed for the detection of cotinine, a secondhand smoke (SHS) biomarker. The devices were fabricated by a craft-cutter to define flow channels, followed by lamination. This approach of cutting/lamination to fabricate LPAD is very similar to making an identification card, offering advantages in simplicity and rugged backing when compared to the common method of patterning paper using SU-8 or wax. We also developed a protocol of localized incision and paper-folding to isolate the detection zone from flow channels; the simple origami step eliminated possible reagent diffusion and flow during antibody immobilization steps and numerous washings. By incorporating luminol-based chemiluminescence for detecting horseradish peroxidase-conjugated cotinine, we employed origami-enabled LPAD to detect cotinine in mouse serum using competitive immunoassay. The detection limit was determined to be 5 ng/mL, a clinically relevant concentration. We believe that LPAD with chemiluminescence detection provides a new platform of low cost and sensitive assays for cotinine detection.