A new device to relieve venipuncture pain can affect haematology test results
Gabriel Lima-Oliveira1, Giuseppe Lippi2, Gian Luca Salvagno3
1Post-Graduate Programme of Pharmaceutical Sciences, Department of Medical Pathology, Federal University of Parana, Curitiba, Parana, Brazil Laboratory of Clinical Biochemistry, Department of Life and Reproduction Sciences, University of Verona, Verona, Italy Clinical Laboratory Bioanalise, Teresina, Piaui, Brazil.
Background:
In vitro diagnostic tests play a key role in patients' management (e.g., guiding red blood cell transfusions). The aim of this study was to evaluate the impact of an innovative device (Buzzy®) which is claimed to be able to relieve venipuncture pain by means of cold and vibration. This device was applied during collection of venous blood by venipuncture for conventional haematology testing.
Materials And Methods:
Blood was drawn from 100 volunteers by a single, expert phlebotomist. A vein was located in the left forearm without applying a tourniquet but using a subcutaneous tissue transilluminator device, so that venous stasis was avoided. Blood samples were collected with a 20G straight needle directly into 4 mL K3EDTA vacuum tubes. In sequence, external cold and vibration was established by Buzzy® on the right forearm -5 cm above the venipuncture site- for 1 minute before venipuncture and continued until the end of the same procedure already performed in the left forearm. Conventional haematological tests were performed using the same instrument (Sysmex® XE-2100D) in all cases.
Results:
When Buzzy® was applied before drawing blood, erythrocyte counts and associated parameters (i.e., haemoglobin and haematocrit) were higher, whereas platelet number, leucocyte count and differential were lower. Statistically and clinically significant differences (P<0.001) were observed for erythrocytes, haemoglobin and haematocrit.
Discussion:
From a practical perspective, cold-induced haemoconcentration promotes the efflux of water, diffusible ions and low molecular weight molecules from the vessel, thus increasing the concentration of other blood analytes at the puncture site. These variations may influence test results, especially for erythrocytes, haemoglobin and haematocrit. The novel Buzzy® device should, therefore, be used with caution when collecting blood for conventional haematological testing because of the observed bias introduced in some parameters.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Blood and Nerve Supply to the Bones
Nutrient Artery
The nutrient artery is the main blood vessel that enters the diaphysis via the nutrient foramen. While most long bones have only one nutrient foramen, large bones, such as the femur, may have two. This...
Local Anesthetics: Clinical Application as Intravenous Regional Anesthesia
One of the advantages of...
Hemodialysis I: Introduction
Errors occurring during blood pressure monitoring
Several factors...