Outcomes of defibrillator therapy in catecholaminergic polymorphic ventricular tachycardia
Ferran Roses-Noguer1, Julian W E Jarman1, Jonathan R Clague1
1Department of Cardiology, Royal Brompton Hospital, London, United Kingdom.
Background:
Catecholaminergic polymorphic ventricular tachycardia (CPVT) is an inherited arrhythmia syndrome characterized by adrenergically induced ventricular arrhythmias in patients with structurally normal hearts. Initiating triggered arrhythmias, such as bidirectional ventricular tachycardia, often degenerate into reentrant arrhythmias, such as ventricular fibrillation (VF).
Objective:
To determine whether the effectiveness of implantable cardioverter-defibrillator (ICD) shocks is dependent on rhythm type.
Methods:
It is a retrospective study of patients with CPVT who had undergone ICD implantation. Thirteen patients received ICDs (median age 15 years; range 9-43 years): 7 of 13 (54%) for cardiac arrest and 6 of 13 (46%) for syncope despite drug therapy. The median follow-up duration was 4.0 years (range 1.7-19.9 years). Nineteen reinterventions occurred, excluding generator replacements. Ten patients received 96 shocks (median 4 shocks; range 1-30 shocks). Eighty-seven shock electrograms were reviewed. Sixty-three (72%) shocks were appropriate, and 24 (28%) were inappropriate (T-wave oversensing 7 [29%], supraventricular arrhythmia 16 [67%], after self-terminating VF 1 [4%]).
Results:
Among appropriate shocks, 20 (32%) were effective in terminating sustained arrhythmia and 43 (68%) were ineffective. Shocks delivered to triggered arrhythmias nearly always failed (1 of 40 [3%] effective), while shocks delivered to VF were usually successful (19 of 23 [83%] effective; P < .001). Four patients received 17 appropriate antitachycardia pacing therapies for ventricular tachycardia: only 2 (12%) were effective. No patient died.
Conclusions:
The effectiveness of ICD shock therapy in CPVT depends on the mechanism of the rhythm treated. Shocks delivered to initiating triggered arrhythmias nearly always fail, whereas those for subsequent VF are usually effective. ICD programming in these patients is exceptionally challenging.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
09:47A Rat Model of Ventricular Fibrillation and Resuscitation by Conventional Closed-chest Technique
Published on: April 26, 2015
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Cardiopulmonary Resuscitation IV: Pharmacological Management
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy V: Interprofessional Care
Cardiopulmonary Resuscitation III: AED Use
Dysrhythmias VI: Management of Dysrhythmias