Related Experiment Video
Updated: May 7, 2026

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
High-resolution MRI of uveal melanoma using a microcoil phased array at 7 T
J W M Beenakker1, G A van Rijn, G P M Luyten
1C. J. Gorter Center for High Field MRI, Department of Radiology, Leiden University Medical Center, Leiden, the Netherlands; Department of Ophthalmology, Leiden University Medical Center, Leiden, the Netherlands.
Abstract:
High-field MRI is a promising technique for the characterisation of ocular tumours, both in vivo and after enucleation. For in vivo imaging at 7 T, a dedicated three-element microcoil array was constructed as a high-sensitivity receive-only device. Using a dedicated blink/fixation protocol, high-resolution in vivo images could be acquired within 3 min in volunteers and patients with no requirement for post-acquisition image registration. Quantitative measures of axial length, aqueous depth and lens thickness in a healthy volunteer were found to agree well with standard ocular biometric techniques. In a patient with uveal melanoma, in vivo MRI gave excellent tumour/aqueous body contrast. Ex vivo imaging of the enucleated eye showed significant heterogeneity within the tumour.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies VII: Vascular Imaging
Imaging Studies IV: Magnetic Resonance Imaging