Assessment of Cardiac Function and Myocardial Morphology Using Small Animal Look-locker Inversion Recovery SALLI MRI in Rats
Published on: July 19, 2013
T1 mapping in the rat myocardium at 7 tesla using a modified CINE inversion recovery sequence
Henk Smit1, Ruben Pellicer Guridi, Jamal Guenoun
1Departments of Medical Informatics and Radiology, Erasmus MC, Rotterdam, The Netherlands.
Purpose:
To evaluate the reproducibility and sensitivity of the modified CINE inversion recovery (mCINE-IR) acquisition on rats for measuring the myocardial T1 at 7 Tesla.
Materials And Methods:
The recently published mCINE-IR acquisition on humans was applied on rats for the first time, enabling the possibility of translational studies with an identical sequence. Simulations were used to study signal evolution and heart rate dependency. Gadolinium phantoms, a heart specimen and a healthy rat were used to study reproducibility. Two cryo-infarcted rats were scanned to measure late gadolinium enhancement (LGE).
Results:
In the phantom reproducibility studies the T1 measurements had a maximum coefficient of variation (COV) of 1.3%. For the in vivo reproducibility the COV was below 5% in the anterior cardiac segments. In simulations with phantoms and specimens, a heart rate dependency of approximately 0.5 ms/bpm was present. The T1 maps of the cryo-infarcted rats showed a clear lowering of T1 in de LGE region.
Conclusion:
The results show that mCINE-IR is highly reproducible and that the sensitivity allows detecting T1 changes in the rat myocardium.