Related Experiment Video
Updated: May 7, 2026

Measurement of Tumor T2* Relaxation Times after Iron Oxide Nanoparticle Administration
Published on: May 19, 2023
R2* as a surrogate measure of ferriscan iron quantification in thalassemia
Wesley C Chan1, Zahra Tejani, Faisal Budhani
1Department of Radiology, Health Sciences Centre, Memorial University, St. John's, NL, Canada.
Purpose:
To determine whether R2* values are a consistent predictor of hepatic iron concentration (HIC) in thalassemia patients by demonstrating a correlation between R2* relaxation rates and FerriScan-determined HIC.
Materials And Methods:
Eighty-eight patients with thalassemia major were retrospectively evaluated. All patients underwent FerriScan imaging and multiecho gradient echo imaging. The results from FerriScan analysis were fitted against R2* estimates using linear regression.
Results:
There was a very strong linear correlation between R2* values and FerriScan-determined HIC (Spearman correlation of 0.976; 95% confidence interval [CI]: 0.963, 0.984).
Conclusion:
R2* values can predict HIC determined by FerriScan using a linear calibration curve. This technique may provide a potentially cost-saving alternative for hepatic iron determination and improve acceptance by referring physicians.