Hemodialysis clearance of iosimenol, a novel iso-osmolar radiographic contrast medium
Yukihiro Hirao1, Eiji Kashiyama1, Ken Umehara1
1Tokushima Research Institute, Otsuka Pharmaceutical Co., Ltd., Tokushima, Japan.
Background:
Iodinated contrast media (CM) have molecular and pharmacokinetic properties likely to make them highly dialyzable. Controlled clinical studies allowing for comparisons of hemodialysis clearance between different test substances and in multiple hemodialysis filters are, however, complex and not always practically feasible. A miniaturized in vitro method was therefore developed to evaluate the dialyzability of a new CM.
Purpose:
To evaluate hemodialysis clearance of iosimenol, a novel iso-osmolar contrast medium (CM), in a select variety of hemodialysis filters and in comparison to commercially available CM.
Material And Methods:
Three different high-flux and one low-flux membrane were used in miniaturized dialyzers to evaluate the in vitro blood clearance of iosimenol. Commercially available CM (iodixanol and iohexol) served as control substances. In vitro dialysis parameters were then used to predict clinical hemodialysis clearances. Residual ratios of endogenous substances (inorganic phosphate, urea nitrogen, creatinine, total bilirubin, and albumin) were used as proof of reliability of the in vitro dialysis system.
Results:
Dialyzable small endogenous molecules were readily eliminated in all membranes. The removal ratios of iosimenol were generally similar to that of iodixanol in all membranes except the high-flux polysulfone but were consistently lower than that of iohexol. The blood clearance of iosimenol during clinical hemodialysis was predicted as, on average, approximately 85 mL/min with the high-flux membranes and 47 mL/min with the low-flux membrane.
Conclusion:
The dialyzability of iosimenol was evaluated using a newly developed in vitro dialysis system, and iosimenol was readily cleared from blood with all four tested membranes. And it is suggested that the dialysis parameters can predict clinical hemodialysis clearance of CM.
More Related Videos
13:21Preparation, Purification, and Characterization of Lanthanide Complexes for Use as Contrast Agents for Magnetic Resonance Imaging
Published on: July 21, 2011
10:50Osmotic Drug Delivery to Ischemic Hindlimbs and Perfusion of Vasculature with Microfil for Micro-Computed Tomography Imaging
Published on: June 29, 2013
Related Concept Videos
Dialysis
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Imaging Studies VII: Vascular Imaging
Renal Drug Clearance: Overview
Renal clearance can be calculated using different methods. One approach is to divide the urinary drug excretion rate by the plasma drug concentration. This method directly measures renal clearance, indicating the kidneys' efficiency in...
Radiological Investigation I: X-ray and CT
Renal Clearance
Renal clearance refers to the volume of plasma cleared of a specific substance, such as creatinine, per unit of time. To measure clearance, urine samples are collected over a 24-hour period during each bladder voiding, followed by a single blood sample at the...