Measurements with the new PHE neutron survey instrument
J S Eakins1, R J Tanner2, L G Hager2
1Public Health England, CRCE, Chilton, Didcot, Oxon OX11 0RQ, UK jonathan.eakins@phe.gov.uk.
Abstract:
A novel design of survey instrument has been developed to accurately estimate ambient dose equivalent from neutrons with energies in the range from thermal to 20 MeV. The device features moderating and attenuating layers to ease measurement of fast and intermediate energy neutrons, combined with guides that channel low-energy neutrons to the single, central detector. A prototype of this device has been constructed and exposed to a set of calibration fields: the resulting measured responses are presented and discussed here, and compared against Monte Carlo data. A simple simulated workplace neutron field has also been developed to test the device.