Measurement of neutron ambient dose equivalent in carbon-ion radiotherapy with an active scanned delivery system
S Yonai1, T Furukawa2, T Inaniwa2
1National Institute of Radiological Sciences, 4-9-1 Anagawa, Inage-ku, Chiba, Japan yonai@nirs.go.jp.
Abstract:
In ion beam radiotherapy, secondary neutrons contribute to an undesired dose outside the target volume, and consequently the increase of secondary cancer risk is a growing concern. In this study, neutron ambient dose equivalents in carbon-ion radiotherapy (CIRT) with an active beam delivery system were measured with a rem meter, WENDI-II, at National Institute of Radiological Sciences. When the same irradiation target was assumed, the measured neutron dose with an active beam was at most ∼15 % of that with a passive beam. This percentage became smaller as larger distances from the iso-centre. Also, when using an active beam delivery system, the neutron dose per treatment dose in CIRT was comparable with that in proton radiotherapy. Finally, it was experimentally demonstrated that the use of an active scanned beam in CIRT can greatly reduce the secondary neutron dose.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Biological Effects of Radiation
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET