Image guidance protocols: balancing imaging parameters against scan time
A M Bates1, J E Scaife, G S J Tudor
1Cancer Research UK VoxTox Research Group, Department of Oncology, University of Cambridge, Addenbrooke's Hospital, Cambridge, UK.
Objective:
Optimisation of imaging protocols is essential to maximise the use of image-guided radiotherapy. This article evaluates the time for daily online imaging with TomoTherapy® (Accuray®, Sunnyvale, CA), separating mechanical scan acquisition from radiographer-led image matching, to estimate the time required for a clinical research study (VoxTox).
Methods:
Over 5 years, 18 533 treatments were recorded for 3 tumour sites of interest (prostate, head and neck and central nervous system). Data were collected for scan length, number of CT slices, slice thickness, scan acquisition time and image matching time.
Results:
The proportion of coarse thickness scans increased over time, with a move of making coarse scans as the default. There was a strong correlation between scan time and scan length. Scan acquisition requires 40 s of processing time. For coarse scans, each additional centimetre requires 8 s for acquisition. Image matching takes approximately 1.5 times as long, so each additional centimetre needs 20 s extra in total. Modest changes to the imaging protocol have minimal impact over the course of the day.
Conclusion:
This work quantified the effect of changes to clinical protocols required for research. The results have been found to be reassuring in the busy National Institutes of Health department.
Advances In Knowledge:
This novel method of data collection and analysis provides evidence of the minimal impact of research on clinical turnover. Whilst the data relate specifically to TomoTherapy, some aspects may apply to other platforms in the future.
More Related Videos
11:49Using High Resolution Computed Tomography to Visualize the Three Dimensional Structure and Function of Plant Vasculature
Published on: April 5, 2013
08:46Using Synchrotron Radiation Microtomography to Investigate Multi-scale Three-dimensional Microelectronic Packages
Published on: April 13, 2016
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT