Anti-Nuclear Antibody Screening Using HEp-2 Cells
Published on: June 23, 2014
Antinuclear antibodies: two-step detection strategy
Maria Roseli Monteiro Callado1, Maria Nancy de Alencar Barroso, Vania Maria Alves
1Hospital Geral de Fortaleza, Rua Ávila Goulart , 900, Fortaleza, 60175295 , Brazil .
Abstract:
In order to evaluate the performance of the chemiluminescent immunoassay (CLIA) in antinuclear antibodies (ANA) testing, using indirect fluorescent antibody (IFA) on HEp-2 cells as a standard, 209 samples were studied from September to October/2010. The tests were conducted according to the procedures recommended by the manufacturers of the reagents. The interpretation of the IFA results was done according to the Brazilian standards. The charts of patients showing different results between the two techniques were analyzed. The CLIA efficiency was 89%, with a sensitivity of 65%, and a specificity of 94.7%. Nine (4.3%) false-positive and 14 (6.7%) false-negative results were detected. Of these, 13 (93%) represented no risk for the diagnosis and therapeutic management of the patients. The CLIA methodology reduced the need for the IFA manual technique by 77% (160/209). The ANA screening test proved to be a fast and acceptable methodology in the studied population. We established the following criteria for the introduction of an automated ANA screening: (1) Positive results must be confirmed by IFA to characterize the pattern and titer of the antibody. (2) Negative results are issued with a notice to request a new test by IFA when the clinical suspicion of autoimmune disease persists.
More Related Videos
10:05Simultaneous Distinction of Monospecific and Mixed DFS70 Patterns During ANA Screening with a Novel HEp-2 ELITE/DFS70 Knockout Substrate
Published on: January 17, 2018
08:26Development and Validation of an Ultrasensitive Single Molecule Array Digital Enzyme-linked Immunosorbent Assay for Human Interferon-α
Published on: June 14, 2018