Objectives:
The purpose of this study was to assess the occurrence, predictors, and mechanisms of optical coherence tomography (OCT)-detected coronary evaginations following drug-eluting stent (DES) implantation.
Background:
Angiographic ectasias and aneurysms in stented segments have been associated with a risk of late stent thrombosis. Using OCT, some stented segments show coronary evaginations reminiscent of ectasias.
Methods:
Evaginations were defined as outward bulges in the luminal contour between struts. They were considered major evaginations (MEs) when extending ≥3 mm along the vessel length, with a depth ≥10% of the stent diameter. A total of 228 patients who had sirolimus (SES)-, paclitaxel-, biolimus-, everolimus (EES)-, or zotarolimus (ZES)-eluting stents implanted in 254 lesions, were analysed after 1, 2, or 5 years; and serial assessment using OCT and intravascular ultrasound (IVUS) was performed post-intervention and after 1 year in 42 patients.
Results:
Major evaginations occurred frequently at all time points in SES (∼26%) and were rarely seen in EES (3%) and ZES (2%, P = 0.003). Sirolimus-eluting stent implantation was the strongest independent predictor of ME [adjusted OR (95% CI) 9.1 (1.1-77.4), P = 0.008]. Malapposed and uncovered struts were more common in lesions with vs. without ME (77 vs. 25%, P < 0.001 and 95 vs. 20%, P < 0.001, respectively) as was thrombus [49 vs. 14%, OR 7.3 (95% CI: 1.7-31.2), P = 0.007]. Post-intervention intra-stent dissection and protrusion of the vessel wall into the lumen were associated with an increased risk of evagination at follow-up [OR (95% CI): 2.9 (1.8-4.9), P < 0.001 and 3.3 (1.6-6.9), P = 0.001, respectively]. In paired IVUS analyses, lesions with ME showed a larger increase in the external elastic membrane area (20% area change) compared with lesions without ME (5% area change, P < 0.001).
Conclusion:
Optical coherence tomography-detected MEs are a specific morphological footprint of early-generation SES and are nearly absent in newer-generation ZES and EES. Evaginations appear to be related to vessel injury at baseline; are associated with positive vessel remodelling; and correlate with uncoverage, malapposition, and thrombus at follow-up.
Related Concept Videos
Coronary Circulation
Coronary circulation begins at the base of the aorta, where two main arteries arise—the left and right coronary arteries. These arteries encircle the heart in the coronary sulcus and supply the...
Coronary Artery Disease I: Introduction
Coronary Artery Disease II: Pathophysiology
Coronary Artery Disease III: Clinical Manifestations
Coronary Artery Disease IV: Preventive Measures
Acute Coronary Syndrome III: Diagnostic Studies