Related Experiment Video
Updated: May 6, 2026

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
A 76-year old man with a torn Freedom SOLO bioprosthesis
Laurens W Wollersheim1, Wilson W Li, Jan van der Meulen
1Division of Cardiothoracic Surgery, Academic Medical Center Amsterdam, Amsterdam, Netherlands.
Abstract:
We describe a case of a 76-year old male who presented with progressive dyspnoea. He underwent an aortic valve replacement with a Freedom SOLO bioprosthesis 6 years ago. Transthoracic echocardiography showed a moderate-to-severe leakage of the Freedom SOLO bioprosthesis. During surgical reintervention, a partial tear of the left coronary cusp was seen from the commissure of the right coronary cusp to its base. After radiographic and microscopic examination, no clear cause was found for the failure of this Freedom SOLO bioprosthesis. To our knowledge, this is the third failure of a Freedom SOLO bioprosthesis reported in the literature. When the long-term follow-up of the Freedom SOLO bioprosthesis is available, it has to be compared with other bioprosthesis for long-term durability.