Related Experiment Video
Updated: May 6, 2026

Removal of Arsenic Using a Cationic Polymer Gel Impregnated with Iron Hydroxide
Published on: June 28, 2019
Method development for the determination of arsenic by sequential injection/anodic stripping voltammetry using
Eakkasit Punrat1, Suchada Chuanuwatanakul, Takashi Kaneta
1Electrochemistry and Optical Spectroscopy Research Unit, Department of Chemistry, Faculty of Science, Chulalongkorn University, Phayathai Road, Pathumwan, Bangkok 10330 Thailand; Graduate School of Natural Science and Technology, Okayama University, Tsushimanaka 1-1-1, Kitaku, Okayama 700-8530, Japan.
Abstract:
An automated method has been developed for determining the concentration of inorganic arsenic. The technique uses sequential injection/anodic stripping voltammetry with a long-lasting gold-modified screen-printed carbon electrode. The long-lasting gold electrode was electrochemically deposited onto a screen-printed carbon electrode at a potential of -0.5 V vs. Ag/AgCl in a supporting electrolyte solution of 1M hydrochloric acid. Under optimal conditions and the applied potentials, the electrode demonstrated that it can be used for a long time without a renewal process. The linear range for the determination of arsenic(III) was 1-100 μg L(-1), and the limit of detection (LOD) in standard solutions was as low as 0.03 μg L(-1) for a deposition time of 120 s and sample volume of 1 mL. This method was used to determine the concentration of arsenic(III) in water samples with satisfactory results. The LOD in real samples was found to be 0.5 μg L(-1). In addition, speciation between arsenic(III) and arsenic(V) has been achieved with the proposed method using deposition potentials of -0.5 V and -1.5 V for the determination of the arsenic(III) concentration and the total arsenic concentration, respectively; the results were acceptable. The proposed method is an automated system that offers a less expensive alternative for determining trace amounts of inorganic arsenic.
More Related Videos
08:21Determination of Inorganic Arsenic in a Wide Range of Food Matrices using Hydride Generation - Atomic Absorption Spectrometry.
Published on: September 1, 2017
10:48Electrochemical Preparation of Poly3,4-Ethylenedioxythiophene Layers on Gold Microelectrodes for Uric Acid-Sensing Applications
Published on: July 28, 2021
Related Concept Videos
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Electrodeposition
Electrodeposition can...
Electrogravimetric Analysis: Overview
To test the completeness of the...
Voltammetric Techniques: Linear-Scan (E vs Time)
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...