Pedometer accuracy during stair climbing and bench stepping exercises.
Makoto Ayabe1, Junichiro Aoki, Kojiro Ishii
1School of Health and Sports Science, Juntendo University Hiragagakuendai , Chiba ; Faculty of Health and Sports Science, Fukuoka University , Fukuoka.
Journal of Sports Science & Medicine
|October 24, 2013
Summary
Pedometer accuracy for stair climbing is reliable for steps taken at 80 per minute or faster on 18 cm or higher surfaces. Certain pedometers (KZ and YM) accurately measure steps during stair climbing and bench stepping exercises.
Area of Science:
- Biomechanics
- Wearable Technology
Background:
- Pedometer use is common for tracking physical activity, but accuracy during activities like stair climbing requires validation.
- Stair climbing and bench stepping involve vertical displacement and varied stepping rates, posing challenges for step-counting devices.
Purpose of the Study:
- To evaluate the accuracy of three commercial pedometers (DW-800, YM, HJ-700IT; OM, Lifecorder; KZ) during stair climbing and bench stepping exercises.
- To determine the influence of stepping rate and platform height on pedometer accuracy.
Main Methods:
- Ten healthy men performed stair climbing/descending and bench stepping exercises (10, 20, 30 cm platforms) wearing three pedometers.
- Stepping rates were controlled at 40, 50, 80, 100, and 120 steps per minute.
Main Results:
- Pedometer scores underestimated actual steps at slower rates and lower platform heights.
- At 80-120 steps/min and 20-30 cm platforms, measurement errors were -3.8 ± 10.8% (KZ), -2.1 ± 9.8% (YM), and -11.0 ± 18.9% (OM).
- KZ and YM pedometers demonstrated acceptable accuracy under specific conditions.
Conclusions:
- The KZ and YM pedometers can accurately assess step counts during stair climbing with 20-30 cm platforms at 80-120 steps/min.
- Pedometer accuracy is acceptable for stair climbing at ≥80 steps/min and ≥18 cm height, within a defined error range.
Related Concept Videos
Errors occurring during blood pressure monitoring
1.6K
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
1.6K
Assessment of blood pressure in brachial artery(two-step method)
2.1K
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
2.1K
Distance Corrections
436
To achieve precise distance measurements, especially in surveying and construction, certain corrections must be applied to account for potential sources of error like the standardization errors, temperature variations, and slope adjustments.Standardization error emerges when measurement equipment undergoes changes, such as wear, repairs, or weather impacts. To address this, surveyors compare the equipment’s readings to a standard. This process identifies any deviation that might lead to...
436