Background:
- Impaired cardiac output (CO) response to exercise is characteristic of chronic heart failure (CHF).
- Practical methods for assessing CO during exercise and defining an impaired response are limited.
- The severity of CO impairment correlates with CHF severity and patient prognosis.
Purpose of the Study:
- To assess the feasibility of non-invasive CO measurement during exercise in CHF patients.
- To define an impaired CO response to exercise.
- To determine if an impaired CO response identifies patients with more severe CHF.
Main Methods:
- Forty-six CHF patients and 13 healthy controls underwent cardiopulmonary exercise testing (CPX).
- A novel, non-invasive bioreactance device was used to measure CO and other hemodynamic parameters during rest and exercise.
- An abnormal CO response was defined as a CO slope falling below 95% of the confidence limits relative to oxygen uptake (VO2).
Main Results:
- The novel bioreactance technique successfully estimated CO during exercise.
- An impaired CO slope was associated with lower peak VO2, lower peak CO, a higher VE/VCO2 slope, and a lower oxygen uptake efficiency slope.
- These findings indicate more severe CHF in patients with an impaired CO response.
Conclusions:
- Non-invasive CO measurement during exercise using bioreactance is feasible in CHF patients.
- An impaired CO response to exercise is a marker of more severe CHF, reduced exercise capacity, and inefficient ventilation.
- This method can aid in identifying high-risk CHF patients and complement CPX assessments.
Abstract:
An impaired cardiac output (CO) response to exercise is a hallmark of chronic heart failure (CHF), and the degree to which CO is impaired is related to the severity of CHF and prognosis. However, practical methods for obtaining cardiac output during exercise are lacking, and what constitutes and impaired response is unclear. Forty six CHF patients and 13 normal subjects underwent cardiopulmonary exercise testing (CPX) while CO and other hemodynamic measurements at rest and during exercise were obtained using a novel, non-invasive, bioreactance device based on assessment of relative phase shifts of electric currents injected across the thorax, heart rate and ventricular ejection time. An abnormal cardiac output response to exercise was defined as achieving ≤ 95% of the confidence limits of the slope of the relationship between CO and oxygen uptake (VO2). An impaired CO slope identified patients with more severe CHF as evidenced by a lower peak VO2, lower peak CO, heightened VE/VCO2 slope, and lower oxygen uptake efficiency slope. CO can be estimated during exercise using a novel bioreactance technique; patients with an impaired response to exercise exhibit reduced exercise capacity and inefficient ventilation typical of more severe CHF. Non- invasive measurement of cardiac performance in response to exercise provides a simple method of identifying patients with more severe CHF and may complement the CPX in identifying CHF patients at high risk. Key pointsNon-invasive measurement of cardiac output during exercise is feasible in patients with heart failure.Impairment in the CO response to exercise identifies heart failure patients with more severe disease, lower exercise capacity and inefficient ventilation.Non-invasive measurement of cardiac performance during exercise has potentially important applications for the functional and prognostic assessment of patients with heart failure.
More Related Videos
07:49Author Spotlight: Investigating HR-Dependent Cardiac Function in Mouse Models Through a Novel Atrial-Pacing Approach
Published on: July 21, 2023
06:10Author Spotlight: Assessment of Cardiac Output Calculation by Thermodilution in Pigs for Effective Perfusion Flow During EVLP
Published on: June 28, 2024
Related Concept Videos
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be...