Peak load resistance of the JuggerKnot™ soft anchor technique compared with other common fixation techniques for
Jason Pui Yin Cheung1, Boris Fung, Wing Yuk Ip
1Department of Orthopaedics and Traumatology, University of Hong Kong Medical Centre, Queen Mary Hospital, Hong Kong.
Introduction:
To identify the strongest peak load resistance among four mallet finger fracture fixation methods (Kirschner wire, pull-out wire, tension-band wiring and the JuggerKnot™ (Biomet) soft anchor fixation).
Methods:
Fixation techniques were assigned among 24 specimens from six cadaveric human hands in a randomized block fashion. Peak load resistance was tested at 30°, 45° and 60° of flexion of the distal interphalangeal joint.
Results:
The mean peak load of tension-band wiring was 67.8 N at 60° of flexion which was most superior. The JuggerKnot™ fixation had mean peak loads of 13.35 N (30°), 22.51 N (45°) and 32.96 N (60°). No complications of implant failure or fragmentation of the dorsal fragment was noted.
Conclusions:
Tension-band wiring was the strongest fixation method but was most prominent on the skin surface as seen in three specimens. The JuggerKnot™ soft anchor fixation had similar peak load resistance as k-wire fixation and pull-out wiring.