Updated: May 6, 2026

Monitoring the Wall Mechanics During Stent Deployment in a Vessel
Published on: May 8, 2012
Longitudinal stent deformation elongating to left main trunk
Kensuke Matsushita1, Kentaro Arakawa, Kiyoshi Hibi
1Division of Cardiology, Fujisawa City Hospital, 2-6-1 Fujisawa, Fujisawa, Kanagawa, 251-8511, Japan, matsuken_22@yahoo.co.jp.
A stent used to treat coronary artery disease unexpectedly elongated into the left main trunk during a procedure. This longitudinal stent elongation may be linked to stent design intended to improve deliverability.
Area of Science:
- Interventional Cardiology
- Medical Device Technology
Background:
- Acute coronary syndrome necessitates prompt intervention.
- Coronary artery stenosis poses a significant risk for cardiovascular events.
Observation:
- During percutaneous coronary intervention for left circumflex artery stenosis, an everolimus-eluting stent was implanted.
- Intravascular ultrasound revealed unexpected longitudinal stent elongation into the left main trunk after post-dilatation.
Findings:
- The implanted stent elongated approximately 2 mm proximally, while the distal edge remained stable.
- A second stent was successfully implanted from the left main trunk to the proximal left anterior descending artery to address the complication.
Implications:
- Stent design, specifically the number of links between hoops, may influence longitudinal behavior and contribute to elongation.
- Understanding and mitigating stent elongation is crucial for optimizing interventional cardiology procedures and patient outcomes.
Related Concept Videos
Deformations in a Transverse Cross Section
As the material stretches, it expands or contracts in orthogonal directions to the load. This phenomenon varies...
Deformations in a Symmetric Member in Bending
When the member is segmented into tiny cubic elements, it is observed that the primary stress...
Plastic Deformations
Plastic Deformations
Normal Strain under Axial Loading
Mitral Stenosis I: Introduction