A numerical survey of motion-induced electric fields experienced by MRI operators
1*Istituto Nazionale di Ricerca Metrologica, Torino, Italy; †Dipartimento Energia, Politecnico di Torino, Italy.
Abstract:
This paper deals with the electric field generated inside the bodies of people moving in proximity to magnetic resonance scanners. Different types of scanners (tubular and open) and various kinds of movements (translation, rotation, and revolution) are analyzed, considering the homogeneous human model proposed in some technical Standards. The computations are performed through the Boundary Element Method, adopting a reference frame attached to the body, which significantly reduces the computational burden. The induced electric fields are evaluated in terms of both spatial distributions and local time evolutions. The possibility of limiting the study to the head without affecting the accuracy of the results is also investigated. Finally, a first attempt to quantify the transient effect of charge separation is proposed.