Measuring the Complete-arch Distortion of an Optical Dental Impression
Published on: May 30, 2019
6.9K
Distortion of bar-phantom image by collimator.
1Columbia Hospital, Milwaukee, Wisconsin, USA.
Summary
Medium-energy collimator use can create distorted scintiphoto images, leading to incorrect conclusions about imaging resolution. Visualizing this distortion clarifies collimator performance differences.
Area of Science:
- Nuclear medicine imaging
- Medical physics
Background:
- Collimators are crucial components in gamma cameras, influencing image resolution and quality.
- Different collimators (low-energy vs. medium-energy) have distinct imaging characteristics.
Purpose of the Study:
- To investigate and illustrate image distortion caused by medium-energy collimators in scintigraphy.
- To address the potential misconception of superior resolution due to collimator-induced distortion.
Main Methods:
- Utilized bar-phantom imaging with both low-energy and medium-energy collimators.
- Employed superposition of transparent films of bar-phantoms and collimators to visualize distortion patterns.
Main Results:
- Observed significant image distortion when using the medium-energy collimator.
- Demonstrated that this distortion can visually suggest better resolution than is actually achieved.
Conclusions:
- The distortion artifacts from medium-energy collimators can be misleading regarding imaging resolution.
- Accurate interpretation of scintiphoto quality requires understanding collimator-specific distortion characteristics.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
9.4K
Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
9.4K
Focusing of Light in the Eye
6.3K
Light rays enter the eye through the cornea, a transparent dome-shaped tissue that is the eye's outermost layer. The cornea bends or refracts, light rays traveling to the pupil. The shape of the cornea determines how much of the light is bent and whether the image will be focused correctly on the retina at the back of the eye. Once the light has passed through both refraction layers, it converges into a single focal point onto a small area. This is where photoreceptors start transforming...
6.3K