Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
Optimization of the parameters for diffusion tensor magnetic resonance imaging data acquisition for breast fiber
Yuan Wang1, Xiao-Peng Zhang1, Yan-Ling Li1
1Key Laboratory of Carcinogenesis and Translational Research (Ministry of Education), Department of Radiology, Peking University Cancer Hospital and Institute, Beijing, China.
Introduction:
Diffusion tensor MRI has emerged as a promising tool for the analysis of the microscopic properties of tissues. Optimizing image acquisition parameters is essential for producing high-quality DTI. This study aimed to optimize the parameters for DTI data acquisition for breast fiber tractography at 1.5 T.
Patients And Methods:
A total of 21 healthy volunteers received breast DTI scanning using an ASSET-based EPI technique operated under different parameters including b value, the number of diffusion gradient directions, and spatial resolution. The images were analyzed for signal-to-noise, signal intensity ratio, mean number and length of reconstructive fiber tracts, and fractional anisotropy value.
Results:
The optimal acquisition parameters at 1.5 T for breast DT-MRI fiber tractography were determined as follows: axial 31 direction, b = 600 seconds per mm(2), matrix 128 × 128 with slice thickness of 3 mm.
Conclusion:
The optimization of data acquisition parameters could improve the quality of breast DT-MRI images and assist fiber tractography at 1.5 T.