Related Experiment Video
Updated: May 6, 2026

Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
[Hypoxic accident during pediatric anesthesia due to an inappropriate setting of respirator]
1Centre hospitalier de Villefranche-de-Rouergue, avenue Caylet, 12200 Villefranche-de-Rouergue, France.
Abstract:
A 3-year-old child was anesthetized for ENT examination and surgery. After induction and tracheal intubation, the patient was ventilated (controlled mode). The respirator screen showed information compatible with a failure of intubation: no expired CO2, no expired flow, no alarm of high pressure limit, and no respiratory chest movement. A fall of SpO2 appeared rapidly which recovered after extubation and manual ventilation through a face mask and reintubation. The expiratory CO2 was present when the patient was ventilated manually and disappeared under controlled ventilation. The increase in the value of the maximal insufflation pressure allowed efficient ventilation with an expiratory CO2 curve and showed high ventilation pressure compatible with a bronchospasm. This case report shows that in case of bronchospasm, if the value of the maximal insufflation pressure is low, this may lead to an erroneous diagnosis of failure of intubation.
More Related Videos
Related Concept Videos
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...
Acute Respiratory Failure-III
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes...
Acute Respiratory Failure-IV