Posterior dislocation of the hinge-post extension in a rotating hinge total knee prosthesis
Givenchy Manzano1, Ran Schwarzkopf
1Department of Orthopaedic Surgery, Joint Replacement Service, University of California Irvine Medical Center, 101 The City Drive South, Pavilion III, Building 29, Orange, CA 92868, USA.
Abstract:
The rotating hinge knee prosthesis is a popular intervention in patients lacking stability with highly constrained total knee arthroplasty. Despite improvements in design, nonmechanical and mechanical complications continue to be a problem. Dislocation of the hinge has been widely described, mainly due to the component fracture. Few reports describe isolated dislocation of the rotating stem. We report a case of isolated disengagement of the rotating hinge mechanism, due to severe flexion gap imbalance, leading to subsequent posterior dislocation of the hinge and anterior knee dislocation, in a patient with a history of multiple total knee arthroplasty revisions. This case suggests the importance of the soft tissue balancing, the adequate patellar tracking, and use of a long cylindrical, minimally tapered rotating stem in hinge arthroplasty to minimize hinge dislocation.
Related Concept Videos
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Bones of the Lower Limb: Femur and Patella
Muscles that Move the Leg
Anterior Compartment
The quadriceps femoris, the most visible muscle of the anterior compartment, is integral for leg extension and thigh flexion. It is formed by merging four distinct muscles — the vastus lateralis, vastus medialis, vastus intermedius, and rectus femoris. The quadriceps tendon, a shared tendon of the four quadriceps muscles, is affixed...